Beta Fulltext view is in preview — article structure may vary. Browse all articles
Contents
Clinical Radiology and Imaging Journal Research Article 9 min read

Entrance-Surface-Dose Evaluation of Patients Examined by Diagnostic X-rays at Alhilal Alahmar Medical Center

Yousif Alghoul A*
* Corresponding author
ISSN: 2640-2343  10.23880/crij-16000188  Received: May 20, 2021  Published: August 20, 2021
  views
 21 references
 6 tables
PDF
Keywords
Patient Entrance Surface Dose (ESD) Diagnostic Medical X-rays
Abstract

Medical Exposures from diagnostic X-rays machines are the most significant artificial source of human exposure to sparsely ionizing radiation. However there is lack of available data and relevant information on the evaluation of patient’s entrance surface dose (ESD) from diagnostic X-Rays. In this research, the Entrance Surface Dose (ESD) is estimated for adult patients who have undertaken diagnosis X-ray examinations at Alhilal Alahmar Medical Center. The ESD has been estimated indirectly by using the X-Rays exposure factors for patients. The results of this research showed that the mean patient entrance surface doses (ESD) were 23.27±7.11 (mGy), 4.31±1.41 (mGy), 3.73±0.32 (mGy), 4.23±3.49 (mGy), 11.95±14.42 (mGy) and 3.79±034 (mGy) respectively for Lumbar Spine (LAT), Cervical Spine (LAT/AP), Chest (PA), Foot (AP/LAT), Knee (AP/LAT), and PNS (AP/ PA/LAT). The mean ESD values estimated are found to be slightly higher than the mean ESD reference values. This confirms the requirement for reducing the patients’ doses to the tolerable levels recommended by many international radiation protection commissions.

Introduction

Medical examinations by diagnostic X-rays are an oldest recognized tool. All patients undergo these examinations benefit from them. However there are unavoidable effects from exposure to X-ray radiation. The ionization nature of the X-rays means that there is a risk of the exposure to this radiation. For example, one type of these effects is called deleterious effects, these effects occur when patients overexposure to these radiation. For this reason, applying radiation protection principles, in medical fields that using the X-rays procedures, is necessary. However, all exposures from diagnostic X-rays must be justified and optimized based on benefit and risk [1, 2, 3]. As it’s well known as the X-ray penetrates biological system it transfers a certain amount of energy, whereas this energy may be risky of unit cells building up these biological system of human body [4]. Internationally, there is increased concern about increasing level of patients exposing to diagnostic X-rays. This concern has been noticed in many published recommendations of the International Committee of Radiation Protection (ICRP) [5, 6, 7]. This concern is also observed in many other reports published by radiation protection organizations and in literature [8].

The almost recommendations are directed to oblige countries to provide restricted radiation dose to patients, whose undertaken X-ray examinations, in radiographic centers and, radiology departments those established at hospitals. One of the major significant factors that applied as an indicator of radiation protection, from deleterious effects, is patient dose. The patient dose specified usually by determining the entrance surface dose (ESD) for patients who exposed to diagnostic X-rays. The entrance surface dose (ESD) or skin entrance dose is generally defined as follows: The radiation absorbed dose by air at the point of intersection of X-ray beam axis with the entrance surface of the patient, which including backscatter radiation [9]. The entrance surface dose (ESD) is one of the basic quantities for measuring the patient dose and, as well for optimizing the given patient radiation dose. This quantity is basic criterion for comparing to the other international reference dose levels which is very significant from the point of view for radiation protection [10]. Many different methods for determining the ESD are established. In common, the ESD can be determined mainly by two techniques: Either by direct measurements of using Thermoluminesent dosimeters (TLD). These dosimeters are usually stacked on the patient’s skin before X-ray exposure. The second way is indirectly through mathematical model calculations. These mathematical model calculations are based on the X-ray machine outputs. The ESD may as well be determined through measurements performed on a phantom beside the data from the patient’s exploration [11, 12, 13].

Particularly, applying the TLDs in measuring the ESD for patients involves time consuming and, it is necessary to use special techniques which may not be available at the most radiographic centers. On the other hand, using ionization chambers to measure the ESDs for patients require applying conversion factors to change the ionization chamber (IC) reading to the absorbed dose. This procedure is a little bit difficult method [14]. Due to the hard accessibility of the TLD or the IC instruments, in addition to the lack of the calibration techniques of these instrumentations. To estimate the ESDs for patients the mathematical technique, which explained on the top of the text, is followed in this research. The aim of this research was to estimate the ESDs for patients exposed to diagnostic X-Rays at medical radiographic centers located in Sebha city.

Materials and Methods

This work was carried out in one of radiographic medical center located in city of Sebha. X-ray machine which used to expose patients is called Toshiba machine. The sample of adult patients exposed with this unit was 87 patients. The minimum inherent filtration in this unit is Aluminum equivalent 2 mm Al/75, X-ray rating up to 150 kvp.

The X-Rays entrance surface dose (ESD) for each patient is estimated mathematically based on the X-Rays exposure factors. These factors are connected by using Chuan and Tsai formula [15]. This formula was applied to estimate ESD for many patients those exposed to diagnostic X-rays in six diagnostic examinations. Chuan and Tsai formula is given as follows:

Where Q is X-Ray charge which represents here the exposure given in (mAs), mm.Al gives minimum inherent filtration Aluminum equivalent, kVp represents X-Ray maximum tube voltage, FSD is focus to skin distance. BSF represents back scattering factor, this factor is roughly equated to one for all investigated examinations. The data obtained was analyzed using professional statistical programmes, which are SigmaPlot and Excel.

The estimated values obtained for the ESD from this research were compared with the international ESD values reported in the literature.

At the beginning the patient’s data such as Age and Gender was first recorded and after that the patient was centered by technician at the right position to be prepared for radiographic. The exposure factors such as peak tube voltage (kvp), X-ray charge (mAs) and focus to surface distance (FSD) were recorded at the time of the examination done. This information was recorded for each patient who undergoing the particular diagnostic procedure.

Results

The research work carried out in Alhilal Alahmar medical center located in Sebha city. One X-ray unit was incorporated in this research work. The results of all obtained data are tabulated below in tables listened from 1 to 6. The results of this work presented in these tables including: gender, patient’s age, peak tube voltage (kvp), exposure factor (Q), focus to surface distance (FSD), examination type, field size, exposure occurrence (this factor gives the number of X-Ray images that taken for each patient at the same diagnostic examination) and entrance-surface-dose (ESD) for all different medical cases being exposed to diagnostic X-rays respectively. The results also included some information related to medical procedures which involving different positions of patients: such as Anterior-Posterior (AP) examination and Lateral (LAT) examination. The other medical procedure included is Posterior-Anterior (PA).The mixed procedures are also considered in this research. In this research, the focus was on six human body parts those exposed to X-rays. These parts include: (i) Lumpier Spine, (ii) Cervical Spine, (iii) Chest, (iv) Foot, (v) Knee, (vi) PNS.

Table 7 indicates the mean values of patient’s exposure factors such as kvp, Q, FSD and ESD and the compression with other studies of the estimated ESD and the established international reference levels of ESD that reported in the literature.

No.GenderAgeKVp (KeV)Q (mAs)FSD (cm)Examination
Type
Field Size
(cm2)
ESD (mGy)Occurrence
1F398516090LAT45´4515.848
2F65851608545´4517.767
3M40851208045´4515.047
4F55851607045´4526.196
5F28851207545´4517.118
6F48861607045´4526.68
7M49851607545´4522.817
8F26851608045´4520.054
9F3784966545´4517.88
10M7284967045´4515.346
11F29831206045´4525.498
12F48831206045´4525.498
13F31862007045´4533.517
14M60851207545´4517.116
15F72852006545´4537.967
16F65862007045´4533.515
17F38871206045´45287

Table 1: Estimated ESD for Lumpier Spine.

No.GenderAgeKVp
(KeV)
Q
(mAs)
FSD
(cm)
Examination
Type
Field Size
(cm2)
ESD
(mGy)
Occurrence
1F298248100LAT45´453.581
2F38804810045´453.411
3F28825110045´453.81
4M38825110045´453.81
5F65834810045´453.671
6M52834810045´453.671
7F43834810045´453.671
8M90844810045´453.761
9F528354100AP45´454.131
10F23838410045´456.421
11F338448100LAT45´453.761
12F46838310045´456.341
13M15808010045´455.681
14F38848410045´456.581
15F383884100AP45´451.341
No.GenderAgeKVp
(KeV)
Q (mAs)FSD
(cm)
Examination
Type
Field Size
(cm2)
ESD
(mGy)
Occurrence
1F478548100PA45´453.81
2F34794810045´453.321
3F47854810045´453.851
4M17834810045´453.671
5M62864810045´453.941
6F55854810045´453.851
7M48854810045´453.851
8M16854810045´453.851
9F45854810045´453.851
10F50844810045´453.761
11F40845110045´453.991
12M20754810045´452.991
13F26845410045´454.231
14M26774810045´453.161
15M40824810045´453.581
16M36855110045´454.091
17M52805110045´453.621

Table 2: Estimated ESD for Cervical Spine.

No.GenderAgeKVp
(KeV)
Q (mAs)FSD
(cm)
Examination
Type
Field Size
(cm2)
ESD
(mGy)
Occurrence
1M48542545LAT45´453.991
2F4349258545´450.921
3F1348258545´450.881
4F74502570AP45´451.411
5F74502565LAT45´451.641
6M3550257545´451.231
7F5258649045´452.951
8F4357494945´457.361
9F4358644945´459.951
10F5055644945´458.951
11M7958495045´457.321

Table 3: Estimated ESD for foot.

No.GenderAgeKVp
(KeV)
Q (mAs)FSD
(cm)
Examination
Type
Field Size
(cm2)
ESD
(mGy)
Occurrence
1F30606425AP LAT45´4540.922
2F24606425AP LAT45´4540.922
3M12596465AP LAT45´455.852
4M45616470AP LAT45´455.392
5M55626470AP LAT45´455.572
6F7574960AP LAT45´454.912
7F63631065AP LAT45´451.042
8F65597260AP LAT45´457.722
9M48596460AP LAT45´456.872
10F40596470AP LAT45´455.042
11F32627265AP LAT45´457.272

Table 4: Estimated ESD for Knee.

No.GenderAgeKVp
(KeV)
Q (mAs)FSD
(cm)
Examination
Type
Field Size
(cm2)
ESD
(mGy)
Occurrence
1F208548100AP45´453.852
2F68448100LAT45´453.761
3M8 m٭7745100LAT45´452.961
4F138548100AP45´453.851
5F488448100PA45´453.764
6F408451100PA45´453.993
7F388354100PA45´454.136
8F288348100PA45´453.676
9F428451100PA45´453.995
10F38048100LAT45´453.412
11F28248100LAT45´453.581
12M148751100PA45´454.284
13F28751100AP45´454.281
14F108051100AP45´453.621
15F38048100LAT45´453.412
16M438551100PA45´454.094
17F318251100PA45´453.81

Table 5: Estimated ESD for PNS. m :٭months

Diag. TypeESD (mGy) This
Study
ESD (mGy)
[15,16]
ESD (mGy)
[17]
ESD (mGy)
[18]
ESD (mGy)
[19]
ESD (mGy)
[20]
ESD (mGy)
[21]
Lumpier Spine (LAT)23.27103.722.612.2510610
Cervical Spine (LAT)4.317.252.4٭٭٭٭
Chest (PA)3.737.430.18٭٭٭٭
Foot4.23٭٭٭٭٭٭
Knee11.95٭٭٭٭٭٭
PNS3.79٭٭٭٭٭٭

Table 6: Comparison and Mean Values for Estimated ESDs of Different Examinations.

Discussion

The aim of this research was strived to evaluate the entrance surface dose (ESD) for patients who go through diagnostic examinations by X-rays. This study was conducted at Alhilal Alahmar medical center. The number of patients included in this study was 87 patients. As it’s show from the tabulated data presented above, the protocol of patients’ exposure at Alhilal Alahmar medical center is higher KvP lower mAs for Cervical Spine (LAT/AP), Chest (PA), Knee (AP/ LAT) and PNS (AP/PA/LAT). The Lumpier Spine examination uses a different protocol which is higher mAs lower KvP.

The mean values of estimated ESD for Lumpier Spine, Cervical Spine, Chest, Knee and PNS are 23.27 (mGy), 4.31 (mGy), 3.73 (mGy), 4.23 (mGy), 11.95 (mGy) and 3.79 (mGy) respectively. When these obtained values of ESD compared with previous studies conducted in this city [16, 17] and international reference levels [18, 19, 20, 21] the findings are that: for Lumpier Spine the estimated ESD is a little bit higher than the international reference level. In comparing this examination with other diagnostic examinations we found that the occurrences of each examination occurring from 4 to 8 times per patient. This procedure is definitely maximizing the ESDs for patients experienced this examination.

For Cervical Spine the estimated ESD is within the acceptable range. While for chest the estimated ESD is also within the acceptable range. The minimum ESD value estimated in this study was for chest (PA) while the maximum ESD estimated is for knee.

Based on this discussion we conclude some points: Firstly, most technicians and medical officers still need to be competent in using X-ray machines at radiographic centers. This point is deduced from the surveys performed at this medical center for few mounts during research work. Secondly, all X-ray tubes should undergo calibration tests frequently to minimize higher exposures for patients examined by these diagnostic X-rays.

Acknowledgement

The researcher would extend his great acknowledgement to technician’s staff working at Radiology department in Alhilal Alahmar Medical Center.

References

  1. Ciraj O, Markovic S, Kosutic D (2003) Patient Dosimetry in Diagnostic Radiology. Nuclear Technology and Radiation Protection.
  2. (1991) 1990 Recommendations of the International Commission on Radiological Protection, ICRP Publication 60, Ann ICRP.
  3. Ciraj O, Markovic S, Kosutic D (2004) Patient Dose from Conventional Diagnostic Radiology Procedures in Serbia and Montenegro. The Journal of Preventive Medicine 12(3-4): 26-34.
  4. Alghoul A, Yasir M (2016) Alternative Mathematical Form for Determining the Effectiveness of High-LET Radiations at Lower Doses Region. Int J Radiol Imaging Technol 2(1): 009.
  5. (2000) Avoidance of Radiation Injuries from Medical Interventional Procedures, ICRP Publication 85, Ann ICRP.
  6. WHO (1982) Quality Assurance in Diagnostic Radiology. World Health Organization.
  7. (1985) CEC, Radiation Protection 33 EUR 9728EN; Office for Official of the EURO Pean Community, Luxembourg.
  8. Aliyu A, Mary Morris, Emmanuel D Langa, Garba Ibrahim, et al. (2021) Assessment of Radiation Dose for Patients during X-ray Procedures in University of Maiduguri Teaching Hospital. Journal of Radiation and Nuclear Applications 6(2): 163-169.
  9. (1994) Code of Safe Practice for the Use of X-Rays in Medical Diagnosis. National Radiation Laboratory, Ministry of Health.
  10. International Commission on Radiation Units and Measurements (1970) Radiation Dosimetry: X-rays Generated at Potentials of 5 to 150 kV. Tech Rep 17, Jr. ICRU 2(15).
  11. Jornet N, Muñoz J, Martin Viera J, Jurado D, Pallerol R, et al. (2013) Determination of Entrance Surface Dose in Standard Explorations in Radiodiagnostic. New York.
  12. Compagnome G, Pagan L, Bergamini C (2005) Comparison of Six Phantoms for Entrance Skin Dose Evaluation in 11 Standard X-ray Examinations. J Appl Clin Med Phys 6(1): 101-113.
  13. Shabon M (2013) Pediatric X-ray Examinations (Special Review). Journal of Nuclear and Radiation Physics 82: 35-44.
  14. Alghoul A (2017) Mathematical Evaluation of Entrance Surface Dose (ESD) for Patients Examined by Diagnostic X-Rays. Open Access J Sci 1(1): 00003.
  15. Ding C, Wang JP, Yu XR, He SS, Yu FH, et al. (2001) Medical Radiation Usage and Exposures from Medical Radiation X-ray. Diagnostic in Shandong Province of China. Radiat Prot Dosimetry 93(2): 261-266.
  16. Chuan J, Tsai H (1999) Evaluation of Gonad Dose and fetal doses for Diagnostic Radiology. Proc Natl Sci Counc Repub China B 23(3): 107-113.
  17. Yousif A, Suasey H A, Ramdan M M (2016) Estimation of Entrance Surface Dose (ESD) of Patients Examined by Diagnostic X-Ray at Sebha Medical Center. Journal of Applied Science 16(1).
  18. Tamboul J, Yousef M, Mokhtar K, Alfaki A, Sulieman A (2014) Assessment of Entrance Surface Dose for the Patients from Common Radiology Examinations in Sudan. Life Science Journal 11(2): 1-5.
  19. (1996) Diagnostic Reference Levels in Medical Imaging, review and additional Advice. a web module produced by committee 3 of the international commission on radiological protection (ICRP).
  20. Hull & East Yorkshire. Hospitals Radiation Protection Service (2005) (Reference Value Proposed as DRL).
  21. (1996) Quality Criteria Reference Doses [EUR96], NA.

Cite this article

BibTeX
APA
RIS
@article{yousif2021,
  title   = {Entrance-Surface-Dose Evaluation of Patients Examined by Diagnostic X-rays at Alhilal Alahmar Medical Center},
  author  = {Yousif Alghoul A},
  journal = {Clinical Radiology and Imaging Journal},
  year    = {2021},
  volume  = {5},
  number  = {2},
  doi     = {10.23880/crij-16000188}
}
Yousif Alghoul A (2021). Entrance-Surface-Dose Evaluation of Patients Examined by Diagnostic X-rays at Alhilal Alahmar Medical Center. Clinical Radiology and Imaging Journal, 5(2). https://doi.org/10.23880/crij-16000188
TY  - JOUR
TI  - Entrance-Surface-Dose Evaluation of Patients Examined by Diagnostic X-rays at Alhilal Alahmar Medical Center
AU  - Yousif Alghoul A
JO  - Clinical Radiology and Imaging Journal
PY  - 2021
VL  - 5
IS  - 2
DO  - 10.23880/crij-16000188
ER  -